Bub­ble Vis­come­ters

 

BYK-Gard­ner bub­ble vis­come­ters are used to quickly deter­mine kine­matic vis­cos­ity of known liquids such as resins and var­nishes.

  • The liquid stan­dards are sealed in glass tubes
  • Shelf life is 15 years
  • Sam­ple tubes can be cleaned quickly and eas­ily
  • Tubes have no ori­fices that can be clogged to cause faulty mea­sure­ments
  • Repeated read­ings may be taken eas­ily once the tem­per­a­ture has been con­trolled

The time required for an air bub­ble to rise is directly pro­por­tio­nal to the vis­cos­ity of the liquid – the faster the bub­ble rises, the lower the vis­cos­ity. BYK-Gard­ner bub­ble vis­come­ters come in let­tered tubes A5 through Z10 in four dif­fer­ent tube sets cov­er­ing vis­cos­ity ranges from 0.05 to 1,000 stokes.

Our bub­ble tubes can be recer­ti­fied to NIST trace­able stan­dards.

Comes com­plete with:
Bub­ble Vis­come­ters:
Bub­ble tube stan­dards
Stor­age case
Two empty tubes
Oper­at­ing man­ual
Cer­ti­fied Sets come with cer­tifi­cate

Bub­ble Vis­come­ter A-T

Note: Indi­vi­d­ual replace­ment tubes can be ordered separately.

Ordering Information Technical Specifications
Cat. No. Description PriceStokesText only
PV-0500 Bubble Viscometer A5-A1 0.05 - 0.31Set of 5 bubble tube standards A5 - A1 with 2 empty tubes (Grade A)
PV-0600 Cert. Bubble Viscometer A5-A1
PV-0510 Bubble Viscometer A-T 0.5 - 5.5Set of 20 bubble tube standards A - T with 2 empty tubes (Grade A)
PV-0610 Cert. Bubble Viscometer A-T
PV-0540 Bubble Viscometer U-Z6 6.66 - 151Set of 12 bubble tube standards U - Z6 with 2 empty tubes (Grade A)
PV-0640 Cert. Bubble Viscometer U-Z6
PV-0560 Bubble Viscometer Z7-Z10 406- 1190Set of 4 bubble tube standards Z7 - Z10 with 2 empty tubes (Grade A)
PV-0660 Cert. Bubble Viscometer Z7-Z10
Standards
AOC Method Ka 6-63
ASTM D 1131, D 1545, D 1725
FTMS 141a Method 4272

Bub­ble Vis­come­ter Pro­ce­dure

  • Know­ing the approx­i­mate vis­cos­ity, pick four stan­dard tubes clos­est in vis­cos­ity to your sam­ple
  • Fill the sam­ple tube with liquid, insert a cork, and then using the tube holder PV-0577, insert the four let­tered tubes and the sam­ple tube into the holder
  • Turn over the holder and visu­ally com­pare what let­ter best matches the rise time of the bub­ble in the sam­ple
  • The rise time in seconds of the sealed tubes and sam­ples can also be deter­mined using a basic timer

Please be aware of the fol­low­ing accu­ra­cies when per­form­ing the test:
Tem­per­a­ture con­trol: 1 °C = 10 % error
Ver­ti­cal­ity con­trol: 5° slant = 10 % error
Tube I.D. con­trol: 0.1 mm = 2 % error

Comes com­plete with:
Empty tubes in lots of 144 per pack­age includ­ing corks

ASTM D 1545 Timer Method
The tube has three amber ring marks at 27, 100 and 108 mm from the bot­tom. Fill the tube up to the 100 mm line, insert the cork down to the 108 mm line and turn the tube bot­tom up. Turn the tube around, start the stop watch when the air bub­ble crosses the 27 mm line and stop when the bub­ble crosses the 100 mm line.

Note: Amber mark­ings in per­ma­nent stain are located on the empty tubes for estab­lish­ing cor­rect bub­ble size.

Ordering Information Accessories
Cat. No. Description Price
PV-0571 Empty Tubes Grade A Inscription GARDNER MT in amber stain; inside diameter is checked for 10.65 ± 0.025 mm
PV-0573 Empty Tubes Grade B Inscription GARDNER BT in amber stain; economical tube for making routine laboratory or factory comparisons; inside diameter 10.75 mm
PV-0575 Empty Tubes Grade N Inscription GARDNER in amber stain; inside diameter is checked for 10.65 ± 0.025 mm; one additional marking at the bottom of the tube for establishing 73 mm bubble path; ASTM D 1545 term: Timer Tubes
PV-0576 Corks For use in retaining samples; used with all grades of tubes; supplied in lots of 150 per bag
PV-0577 Holder for 5 Tubes Standards and samples are placed parallel to each other in a true vertical position; sturdy metal frame with plastic handle; fits up to 5 tubes; the flat area allows the holder to sit in a water bath or on a lab bench; comes without tubes