Training spectro-guide
BYK-Gardner offers you more than just an instrument. We assist you in analyzing your color readings, understanding how to set tolerances and as a result be able to use the spectro-guide to save time and money and at the same time improve your quality. Therefore, the instrument comes with a half-day training course including:

1. Color Theory
• The building blocks of color: illuminant, observer, object
• Color differences with interpretation

2. Operation and Software Training
• Measure samples and standards by single and average readings
• Save, recall and delete measurements
• Change illuminants, observers, color scales
• Direct data transfer to easy-link

Comes complete with:
spectro-guide
Black calibration standard
White calibration standard with certificate
Green checking reference
High gloss standard
Sample area locator
Software easy-link
Interface cable
4 x AA batteries
Hand strap
Carrying case
Operating instructions
Color theory folder
Training

For Calibration Services see page 331 - 335.

¹Gloss units

Gloss
Color
Standards
ColorGloss
ASTMD 2244, E 308, E 1164D 523, D 2457
DIN5033, 5036, 617467530
ISO77242813, 7668
Ordering Information Technical Specifications
Cat. No. Description Color Geometry Gloss Geometry Color Aperture Gloss Aperture
CC-6801 spectro-guide 45/0 gloss 45/0 60° 11 mm 5 x 10 mm
CC-6802 spectro-guide 45/0 gloss S 45/0 60° 11 mm 5 x 10 mm
CD-6834 spectro-guide sphere gloss d/8 spin 60° 11 mm 5 x 10 mm
CD-6836 spectro-guide sphere gloss S d/8 spin 60° 11 mm 5 x 10 mm
SE-6800 Extended Warranty one year additional
Memory 1500 Standards, 999 Samples
Languages English; German; French; Italian; Spanish; Japanese; Chinese
Power Supply 4 AA alkaline; NiCd or MH batteries
Operating Temperature 50 °F - 110 °F (10 °C - 42 °C)
Humidity < 85% relative humidity, non-condensing / 95 °F
(35 °C)
Dimensions
(H x W x D)
3.7 x 3.2 x 7 in (9.5 x 8 x 18 cm)
Weight approx. 1.1 lbs (approx. 0.5 kg)
0 to 100 GU S family (0 to 10 GU)
Repeatability ±0.2 GU¹ ±0.1 GU¹
Reproducibility ±1.0 GU¹ ±0.5 GU¹
Spectral Range 400 - 700 nm, 10 nm resolution
Repeatability 0.01 ΔE*, 1 ơ
(10 consecutive measurements on white)
Reproducibility 0.2 ΔE*, 1 ơ (average on 12 BCRA II tiles)
Color Systems CIELab/Ch; Lab(h); XYZ; Yxy
Color Differences ΔE*; ΔE(h); ΔEFMC2; ΔE94; ΔECMC; ΔE99; ΔE2000
Indices YIE313; YID1925; WIE313; CIE; Berger; Color strength; Opacity; Metamerism
Illuminants A; C; D50; D55; D65; D75; F2; F6; F7; F8; F10; F11; UL30
Observer 2°; 10°